Command: /arckit:uk-mdr-classification | Overlay: UK NHS Clinical Safety Overlay | Origin: Community-contributed

Purpose

Determines whether a software product is a medical device under UK MDR 2002 (as amended) and EU MDR 2017/745, classifies it (Class I / IIa / IIb / III), identifies the marking pathway (UKCA / UKNI / CE), the conformity-assessment route (self-declaration / Approved Body / Notified Body), the QMS expectation, the standards alignment expected, and the post-market obligations that flow from the classification.

The output is not regulatory advice. Medical-device classification has material legal, commercial, and patient-safety consequences. The artefact MUST be reviewed and signed off by a qualified Regulatory Affairs Specialist or Notified Body / Approved Body advisor.

Scope

Software-as-medical-device (SaMD) and AI-as-medical-device (AIaMD) only. Hardware medical devices and software embedded in hardware devices are out of scope for v1 — different conformity routes, different test regimes, different MHRA processes. Consult specialist regulatory advice for hardware.

What it produces

Single artefact projects/{NNN}-<slug>/ARC-{NNN}-NHSMDR-v1.0.md covering:

  • Intended-purpose statement captured verbatim from upstream artefacts (the load-bearing input — small wording changes change the classification)
  • Scope determination — UK MDR 2002 regulation 2 test, MHRA Stand-Alone Software decision tree, MHRA Borderline Manual citations. Explicit "Not a Medical Device" determination supported (and often the right answer for record-keeping, signposting, appointment booking, and similar tools)
  • UK MDR 2002 classification — Class I / IIa / IIb / III, subclass flags, self-certification eligibility
  • EU MDR 2017/745 classification — Rule 11 reasoning, divergence from UK classification (EU is typically more conservative; many products that were Class I under MDD moved to Class IIa under MDR)
  • Marking pathway — UKCA (GB), UKNI (NI under Windsor Framework), CE (EU market + NI), CE-in-GB transitional arrangements
  • Conformity-assessment route — self-declaration vs Approved Body vs Notified Body, with technical documentation and clinical evaluation expectations
  • MHRA SaMD / AIaMD Programme alignment — applicable work packages from the MHRA Software and AI as a Medical Device Change Programme roadmap (WP1 Qualification, WP2 Classification, WP3 Premarket requirements, WP4 Post Market, WP5 Cyber Secure Medical Devices; for AIaMD also WP9 AI Rigour, WP10 Project Glass Box, WP11 Project Ship of Theseus)
  • Standards alignment — ISO 14971 (risk management), IEC 62304 (software lifecycle), ISO 13485 (QMS, signposted), IEC 62366-1 (usability), ISO/IEC 27001 (information security), ISO/IEC TR 24028 (AI trustworthiness for AIaMD)
  • Post-market obligations — PMS plan, vigilance reporting timelines, PSUR cadence, FSCA, AIaMD substantial-change handling
  • Substantial change triggers, open regulatory risks

When to use

  • Any digital health product where there is uncertainty about whether it is a medical device
  • Clinical decision-support tools, monitoring tools, diagnostic tools, triage tools, risk-score tools
  • AI/ML products with any clinically-facing function
  • Products placed on the NI market (where EU MDR applies under the Windsor Framework)

Prerequisites

Requires the project's REQ (especially the intended-purpose statement), STKE, DATA, and DCB0129 case (if present — clinical hazards inform the ISO 14971 risk file expected under both UK MDR and EU MDR).

"Not a medical device" is a defensible answer

Many digital-health products are not medical devices — record-keeping tools, appointment booking, signposting, navigation. The MHRA Borderline Manual is the canonical source for these determinations; the artefact cites the closest borderline example explicitly when reaching this conclusion. The "Not a Medical Device" path requires the responsible person at the manufacturer to sign off the determination and record that the product will not be marketed as a medical device.

EU MDR Rule 11 is more restrictive than legacy UK readings

Products historically self-certified as Class I under MDD have widely moved to Class IIa under EU MDR. Do not anchor on legacy classifications — the command applies Rule 11 fresh.

MHRA reform of UK MDR

UK MDR is being substantially reformed by MHRA. CE-marking recognition in GB transitional arrangements have moved repeatedly. The artefact pins its assessment date and lists known regulatory uncertainties. Re-run on material MHRA publications.

  • /arckit:uk-nhs-dcb0129 — DCB0129 hazard log feeds the ISO 14971 risk file (complementary; ISO 14971 and DCB0129 are distinct but cross-referenced)
  • /arckit:uk-nhs-dtac — DTAC consumes the classification statement to determine procurement assurance route
  • /arckit:risk — project risk register cross-references regulatory-classification risks (misclassification, market-access risk, post-market surveillance failure)
  • /arckit:adr — major regulatory routing decisions (UKCA self-declare vs Approved Body, EU MDR via Notified Body vs not, NI placement strategy) belong in an ADR
  • /arckit:atrs — if AIaMD, ATRS publication is expected alongside model card